Recall Z-0440-2013— Cooper Surgical, Inc.

Class II · Terminated

Class II FDA medical device recall by Cooper Surgical, Inc., initiated 2012-10-31, terminated 2014-03-04. It names one FDA 510(k) clearance: K932115. Product: Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy. Reason: Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle.

Recalling firm
Cooper Surgical, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-10-31
Terminated
2014-03-04
Firm location
Trumbull, CT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy

Reason for recall

Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle