Recall Z-0437-2013— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2012-09-21, terminated 2014-04-17. It names one FDA 510(k) clearance: K872379. Product: NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.. Reason: Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset.

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-09-21
Terminated
2014-04-17
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.

Reason for recall

Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset