Recall Z-0433-2014— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2013-05-10, terminated 2014-08-26. FDA's recall record names no specific 510(k) clearance for this event. Product: syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.. Reason: Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-05-10
Terminated
2014-08-26
Firm location
Newark, DE, United States

Product description

syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.

Reason for recall

Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

Data sourced from openFDA. This site is unofficial and independent of the FDA.