Recall Z-0430-2015— Intuitive Surgical, Inc.

Class II · Terminated

Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2014-11-03, terminated 2015-02-03. It names one FDA 510(k) clearance: K131861. Product: 8MM,LONG TIP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.. Reason: Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument..

Recalling firm
Intuitive Surgical, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-11-03
Terminated
2015-02-03
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

8MM,LONG TIP FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Reason for recall

Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.

Data sourced from openFDA. This site is unofficial and independent of the FDA.