Recall Z-0429-2016— bioMerieux, Inc.

Class II · Terminated

Class II FDA medical device recall by bioMerieux, Inc., initiated 2015-11-03, terminated 2017-12-04. FDA's recall record names no specific 510(k) clearance for this event. Product: MYLA MASTER DVD V4.1 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).. Reason: MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID..

Recalling firm
bioMerieux, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-11-03
Terminated
2017-12-04
Firm location
Durham, NC, United States

Product description

MYLA MASTER DVD V4.1 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Reason for recall

MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.

Data sourced from openFDA. This site is unofficial and independent of the FDA.