Recall Z-0428-2026— Beckman Coulter, Inc.
Class II · Ongoing
Class II FDA medical device recall by Beckman Coulter, Inc., initiated 2025-10-01. It names one FDA 510(k) clearance: K121790. Product: Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer. Reason: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results..
- Recalling firm
- Beckman Coulter, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-10-01
- Firm location
- Chaska, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Reason for recall
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.