Recall Z-0425-2014— B. Braun Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by B. Braun Medical, Inc., initiated 2013-04-04, terminated 2014-08-26. It names 3 FDA 510(k) clearances: K090905, K904855, K900865. Product: 13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.. Reason: Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i.

Recalling firm
B. Braun Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-04-04
Terminated
2014-08-26
Firm location
Allentown, PA, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Reason for recall

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i