Recall Z-0425-2014— B. Braun Medical, Inc.
Class II · Terminated
Class II FDA medical device recall by B. Braun Medical, Inc., initiated 2013-04-04, terminated 2014-08-26. It names 3 FDA 510(k) clearances: K090905, K904855, K900865. Product: 13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.. Reason: Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i.
- Recalling firm
- B. Braun Medical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-04-04
- Terminated
- 2014-08-26
- Firm location
- Allentown, PA, United States
510(k) clearances named in this recall
3 on FDA's record- K090905 — ADDEASE 20MM BINARY CONNECTOR WITH 17 GA. NEEDLE
- K904855 — ADD-A-VIAL
- K900865 — ADD-A-VIAL II BINARY CONNECTOR
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.
Reason for recall
Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i
Data sourced from openFDA. This site is unofficial and independent of the FDA.