Recall Z-0423-2023— Insulet Corporation

Class II · Ongoing

Class II FDA medical device recall by Insulet Corporation, initiated 2022-11-14. It names one FDA 510(k) clearance: K203768. Product: Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM. Reason: The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan..

Recalling firm
Insulet Corporation
Classification
Class II
Status
Ongoing
Initiated
2022-11-14
Firm location
Acton, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

Reason for recall

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.