Recall Z-0421-2016— Verathon, Inc.

Class II · Terminated

Class II FDA medical device recall by Verathon, Inc., initiated 2015-11-30, terminated 2016-08-22. FDA's recall record names no specific 510(k) clearance for this event. Product: GlideScope AVL Video Laryngoscope (AVL), Part Number AVL 3- 0574-0115; AVL 4-0574-0116, AVL 5- 0574-0117. AVL is a video laryngoscope that incorporates a high resolution color camera, an LED light source and video output to a remote monitor. This system has a digital platform.. Reason: Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine inspection before or after intubation..

Recalling firm
Verathon, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-11-30
Terminated
2016-08-22
Firm location
Bothell, WA, United States

Product description

GlideScope AVL Video Laryngoscope (AVL), Part Number AVL 3- 0574-0115; AVL 4-0574-0116, AVL 5- 0574-0117. AVL is a video laryngoscope that incorporates a high resolution color camera, an LED light source and video output to a remote monitor. This system has a digital platform.

Reason for recall

Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine inspection before or after intubation.