Recall Z-0418-2016— Fujifilm Medical Systems U.S.A., Inc.

Class II · Terminated

Class II FDA medical device recall by Fujifilm Medical Systems U.S.A., Inc., initiated 2015-08-31, terminated 2018-03-21. It names one FDA 510(k) clearance: K944759. Product: Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT, ED-310XU, ED-420XL, ED-410XT, ED-410XU, ED-450XL, and ED-450XT. Medical endoscope for the duodenum and upper G.I. tract. Open elevator wire channel duodenoscopes.. Reason: A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates reprocessing..

Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Classification
Class II
Status
Terminated
Initiated
2015-08-31
Terminated
2018-03-21
Firm location
Wayne, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Fujifilm Medical Systems Duodenoscope, Model Numbers ED-200XU, ED-200XT, ED-250XT, ED-310XU, ED-420XL, ED-410XT, ED-410XU, ED-450XL, and ED-450XT. Medical endoscope for the duodenum and upper G.I. tract. Open elevator wire channel duodenoscopes.

Reason for recall

A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates reprocessing.