Recall Z-0411-2018— Baxter Healthcare Corporation
Class II · Terminated
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2017-12-21, terminated 2020-08-04. It names one FDA 510(k) clearance: K133807. Product: MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541. Reason: Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-12-21
- Terminated
- 2020-08-04
- Firm location
- Deerfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Reason for recall
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.