Recall Z-0409-2013— KCI USA, Inc.

Class II · Terminated

Class II FDA medical device recall by KCI USA, Inc., initiated 2012-06-27, terminated 2013-04-22. FDA's recall record names no specific 510(k) clearance for this event. Product: RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction. Reason: The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers..

Recalling firm
KCI USA, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-06-27
Terminated
2013-04-22
Firm location
San Antonio, TX, United States

Product description

RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction

Reason for recall

The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.