Recall Z-0408-2024— EMD Millipore Corporation

Class II · Ongoing

Class II FDA medical device recall by EMD Millipore Corporation, initiated 2023-10-17. FDA's recall record names no specific 510(k) clearance for this event. Product: Lugol s Iodine Solution 500ML, Part Number 624-71. Reason: Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles..

Recalling firm
EMD Millipore Corporation
Classification
Class II
Status
Ongoing
Initiated
2023-10-17
Firm location
Burlington, MA, United States

Product description

Lugol s Iodine Solution 500ML, Part Number 624-71

Reason for recall

Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles.