Recall Z-0404-2019— Medtronic Navigation, Inc.-Littleton

Class II · Terminated

Class II FDA medical device recall by Medtronic Navigation, Inc.-Littleton, initiated 2018-09-17, terminated 2020-05-28. It names one FDA 510(k) clearance: K173664. Product: Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R. Reason: O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit.

Recalling firm
Medtronic Navigation, Inc.-Littleton
Classification
Class II
Status
Terminated
Initiated
2018-09-17
Terminated
2020-05-28
Firm location
Littleton, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R

Reason for recall

O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit