Recall Z-0402-2018— Philips Medical Systems (Cleveland) Inc

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2017-07-11, terminated 2019-07-02. It names one FDA 510(k) clearance: K060937. Product: Brilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems. Reason: Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions..

Recalling firm
Philips Medical Systems (Cleveland) Inc
Classification
Class II
Status
Terminated
Initiated
2017-07-11
Terminated
2019-07-02
Firm location
Cleveland, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brilliance iCT SP - Model 728311 Brilliance iCT - Model 728306 Computed Tomography X-ray systems

Reason for recall

Numerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.