Recall Z-0401-2017— Siemens Healthcare Diagnostics, Inc.
Class III · Terminated
Class III FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2016-08-25, terminated 2017-12-01. It names one FDA 510(k) clearance: K060264. Product: Dimension Vista IRON Flex reagent cartridges. Reason: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU)..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2016-08-25
- Terminated
- 2017-12-01
- Firm location
- Newark, DE, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dimension Vista IRON Flex reagent cartridges
Reason for recall
Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).
Data sourced from openFDA. This site is unofficial and independent of the FDA.