Recall Z-0399-2022— Beckman Coulter Inc.
Class II · Ongoing
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2021-11-10. It names one FDA 510(k) clearance: K971505. Product: SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825. Reason: Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-11-10
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
Reason for recall
Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
Data sourced from openFDA. This site is unofficial and independent of the FDA.