Recall Z-0397-2018— Smith & Nephew, Inc.
Class II · Terminated
Class II FDA medical device recall by Smith & Nephew, Inc., initiated 2017-12-11, terminated 2018-06-22. It names one FDA 510(k) clearance: K132886. Product: smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.. Reason: Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa.
- Recalling firm
- Smith & Nephew, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-12-11
- Terminated
- 2018-06-22
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Reason for recall
Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
Data sourced from openFDA. This site is unofficial and independent of the FDA.