Class I FDA medical device recall by Customed, Inc, initiated 2014-10-08, terminated 2017-08-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Cesarean Pack, Catalog # 900-660 Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.. Reason: Significant compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection..

Recalling firm
Customed, Inc
Classification
Class I
Status
Terminated
Initiated
2014-10-08
Terminated
2017-08-14
Firm location
Fajardo, PR, United States

Product description

Cesarean Pack, Catalog # 900-660 Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Reason for recall

Significant compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.