Recall Z-0396-2017— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-10-04, terminated 2017-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160. Reason: .Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-10-04
Terminated
2017-06-05
Firm location
Indianapolis, IN, United States

Product description

COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160

Reason for recall

.Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.