Recall Z-0395-2017— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-10-04, terminated 2017-06-05. It names one FDA 510(k) clearance: K072714. Product: Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190. Reason: Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-10-04
Terminated
2017-06-05
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

Reason for recall

Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded.

Data sourced from openFDA. This site is unofficial and independent of the FDA.