Recall Z-0384-2021— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2020-09-24, terminated 2022-05-20. FDA's recall record names no specific 510(k) clearance for this event. Product: ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-26. Reason: Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-09-24
Terminated
2022-05-20
Firm location
Warsaw, IN, United States

Product description

ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-26

Reason for recall

Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.