Recall Z-0383-2022— Medtronic Navigation, Inc.

Class I · Ongoing

Class I FDA medical device recall by Medtronic Navigation, Inc., initiated 2021-11-11. It names 2 FDA 510(k) clearances: K190672, K150216. Product: Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585. Reason: Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury..

Recalling firm
Medtronic Navigation, Inc.
Classification
Class I
Status
Ongoing
Initiated
2021-11-11
Firm location
Louisville, CO, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585

Reason for recall

Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.