Recall Z-0380-2018— RAYSEARCH LABORATORIES AB

Class II · Terminated

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2017-07-17, terminated 2018-08-29. It names one FDA 510(k) clearance: K171536. Product: RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.. Reason: For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Terminated
Initiated
2017-07-17
Terminated
2018-08-29
Firm location
Stockholm, N/A, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

Reason for recall

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.