Recall Z-0379-2017— Abbott Laboratories
Class II · Terminated
Class II FDA medical device recall by Abbott Laboratories, initiated 2016-10-11, terminated 2017-05-17. FDA's recall record names no specific 510(k) clearance for this event. Product: ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.. Reason: Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21)..
- Recalling firm
- Abbott Laboratories
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-10-11
- Terminated
- 2017-05-17
- Firm location
- Abbott Park, IL, United States
Product description
ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.
Reason for recall
Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).