Recall Z-0366-2023— Full Vision Inc

Class II · Ongoing

Class II FDA medical device recall by Full Vision Inc, initiated 2022-08-31. FDA's recall record names no specific 510(k) clearance for this event. Product: FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928. Reason: This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop..

Recalling firm
Full Vision Inc
Classification
Class II
Status
Ongoing
Initiated
2022-08-31
Firm location
Newton, KS, United States

Product description

FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928

Reason for recall

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Data sourced from openFDA. This site is unofficial and independent of the FDA.