Recall Z-0364-2024— SunMed Holdings, LLC
Class II · Ongoing
Class II FDA medical device recall by SunMed Holdings, LLC, initiated 2023-10-11. It names one FDA 510(k) clearance: K012842. Product: Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator REF #s: BVMB812, BVMB812-F, BVMB810, BVMB810-F, BVMB810S-F.. Reason: A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing..
- Recalling firm
- SunMed Holdings, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-10-11
- Firm location
- Grand Rapids, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Ventlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator REF #s: BVMB812, BVMB812-F, BVMB810, BVMB810-F, BVMB810S-F.
Reason for recall
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Data sourced from openFDA. This site is unofficial and independent of the FDA.