Recall Z-0364-2022— Smith & Nephew, Inc.
Class II · Ongoing
Class II FDA medical device recall by Smith & Nephew, Inc., initiated 2021-11-01. It names one FDA 510(k) clearance: K201918. Product: smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201. Reason: Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System..
- Recalling firm
- Smith & Nephew, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-11-01
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Reason for recall
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
Data sourced from openFDA. This site is unofficial and independent of the FDA.