Recall Z-0361-2023— Cardinal Health
Class II · Ongoing
Class II FDA medical device recall by Cardinal Health, initiated 2022-11-01. It names one FDA 510(k) clearance: K903340. Product: Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.. Reason: Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention.
- Recalling firm
- Cardinal Health
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-11-01
- Firm location
- Mansfield, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.
Reason for recall
Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention