Recall Z-0359-2016— ConMed Corporation
Class II · Terminated
Class II FDA medical device recall by ConMed Corporation, initiated 2015-11-05, terminated 2018-05-24. It names one FDA 510(k) clearance: K140459. Product: Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.. Reason: Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle..
- Recalling firm
- ConMed Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-11-05
- Terminated
- 2018-05-24
- Firm location
- Utica, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.
Reason for recall
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.