Recall Z-0357-2019— Beckman Coulter Inc.
Class III · Terminated
Class III FDA medical device recall by Beckman Coulter Inc., initiated 2018-05-24, terminated 2022-12-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Software Package, iPAW 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95343, Manufactured by: Beckman Coulter Biomedical GmbH Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.. Reason: Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2018-05-24
- Terminated
- 2022-12-08
- Firm location
- Brea, CA, United States
Product description
Software Package, iPAW 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95343, Manufactured by: Beckman Coulter Biomedical GmbH Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.
Reason for recall
Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.
Data sourced from openFDA. This site is unofficial and independent of the FDA.