Recall Z-0354-2014— Restoration Robotics

Class II · Terminated

Class II FDA medical device recall by Restoration Robotics, initiated 2013-10-13, terminated 2013-12-10. It names one FDA 510(k) clearance: K123548. Product: ARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; Dissection Punch 4.0mm FP-23304; Saline Nose Cone FP-45387; Follicle Trap - FP-23306 and Non-sterile Disposable Clinical Kit FP -23267): and ARTAS Disposable Kits, 18 Gauge Epsilon (Contains Disposable Kit, 18 Gauge Epsilon FP 45547 and Non-Sterile Disposable Kit FP-23267). RX. Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.. Reason: Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine..

Recalling firm
Restoration Robotics
Classification
Class II
Status
Terminated
Initiated
2013-10-13
Terminated
2013-12-10
Firm location
San Jose, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; Dissection Punch 4.0mm FP-23304; Saline Nose Cone FP-45387; Follicle Trap - FP-23306 and Non-sterile Disposable Clinical Kit FP -23267): and ARTAS Disposable Kits, 18 Gauge Epsilon (Contains Disposable Kit, 18 Gauge Epsilon FP 45547 and Non-Sterile Disposable Kit FP-23267). RX. Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Reason for recall

Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.