Recall Z-0353-2018— Zoll Manufacturing Corp.

Class II · Terminated

Class II FDA medical device recall by Zoll Manufacturing Corp., initiated 2017-09-12, terminated 2021-08-10. FDA's recall record names no specific 510(k) clearance for this event. Product: LifeVest Wearable Defibrillator Model 4000, Product Number 10A0988-A01. The LifeVest system is indicated for patients 18 years of age and older who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.. Reason: Incorrect service code for properly catching critical defects during self-check. Potential for defibrillation shock failure.

Recalling firm
Zoll Manufacturing Corp.
Classification
Class II
Status
Terminated
Initiated
2017-09-12
Terminated
2021-08-10
Firm location
Pittsburgh, PA, United States

Product description

LifeVest Wearable Defibrillator Model 4000, Product Number 10A0988-A01. The LifeVest system is indicated for patients 18 years of age and older who are at risk for sudden cardiac arrest and are not candidates for or refuse an implantable defibrillator.

Reason for recall

Incorrect service code for properly catching critical defects during self-check. Potential for defibrillation shock failure