Recall Z-0349-2024— Becton Dickinson & Company
Class II · Ongoing
Class II FDA medical device recall by Becton Dickinson & Company, initiated 2023-10-11. It names one FDA 510(k) clearance: K022209. Product: BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM Pump Infusion Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 10013072; BD AlarisTM Bag Access Non-Vented Back Check Valve 2 SmartSiteTM Y-Sites, Catalog # 10013361; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 10013890; BD AlarisTM Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing SmartSiteTM Port (Burette) SmartSiteTM Y-Site, Catalog # 10015012; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE-Lined), Catalog # 10015862; BD AlarisTM Pump Infusion Blood Set 180 Micron Filter 2 Blunt Spikes, Catalog # 10062818; BD AlarisTM Pump Infusion Buerette Set SmartSiteTM Port (Burette) 3 SmartSiteTM Y-Sites, Catalog # 10821753; BD AlarisTM Pump Infusion Set 2 Back Check Valves 3 SmartSiteTM Y-Sites, Catalog # 11171447; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 11426964; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Back Check Valve Low Sorbing Tubing (PE Lined) 2 SmartSiteTM Y-sites, Catalog # 11532269; BD AlarisTM Pump Infusion Buerette Set 0.2Micron Filter Ball Valve SmartSiteTM Port (Burette) 2 SmartSiteTM Y-Sites, Catalog # 11613191; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 10015861A; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM ClosedMale LuerWith PrimingCAP, Catalog # 22000-B007T; BD AlarisTM Pump Infusion Set 1.2 Micron Filter, Catalog # 2202-0007; BD AlarisTM Pump Infusion Set, Catalog # 2204-0007; BD AlarisTM Pump Infusion Set Yellow-Striped Microbore Tubing, Catalog # 2206-0007; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 2260-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22600-0007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22602-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22603-B007T; BD AlarisTM Pump Infusion Set Back Check Valve SmartSiteTM Y-site, Catalog # 24001-0007; BD AlarisTM Pump Infusion Set Half Set SmartSiteTM Y-site, Catalog # 2403-0007; BD AlarisTMPumpInfusionSetSmartsiteTM BagAcessNon-Vented0.2MicronFilterBondedTexiumTMClosedLowSorbingTubing(PELined)MaleLuerwithprimingcapSmartSiteTMY-site, Catalog # 24301-0007T; BD AlarisTM Pump Infusion Buerette Set Smallbore Tubing SmartsiteTM Port (Burette) 3 SmartSiteTM Y-Site, Catalog # 2441-0007; BD AlarisTM Pump Infusion Set Back Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSiteTM Y-sites, Catalog # 2450-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 24600-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing TexiumTM Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 24601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-Site, Catalog # 2465-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined), Catalog # 2466-0007.. Reason: Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP)..
- Recalling firm
- Becton Dickinson & Company
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-10-11
- Firm location
- Franklin Lakes, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM Pump Infusion Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 10013072; BD AlarisTM Bag Access Non-Vented Back Check Valve 2 SmartSiteTM Y-Sites, Catalog # 10013361; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 10013890; BD AlarisTM Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing SmartSiteTM Port (Burette) SmartSiteTM Y-Site, Catalog # 10015012; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE-Lined), Catalog # 10015862; BD AlarisTM Pump Infusion Blood Set 180 Micron Filter 2 Blunt Spikes, Catalog # 10062818; BD AlarisTM Pump Infusion Buerette Set SmartSiteTM Port (Burette) 3 SmartSiteTM Y-Sites, Catalog # 10821753; BD AlarisTM Pump Infusion Set 2 Back Check Valves 3 SmartSiteTM Y-Sites, Catalog # 11171447; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 11426964; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Back Check Valve Low Sorbing Tubing (PE Lined) 2 SmartSiteTM Y-sites, Catalog # 11532269; BD AlarisTM Pump Infusion Buerette Set 0.2Micron Filter Ball Valve SmartSiteTM Port (Burette) 2 SmartSiteTM Y-Sites, Catalog # 11613191; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 10015861A; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM ClosedMale LuerWith PrimingCAP, Catalog # 22000-B007T; BD AlarisTM Pump Infusion Set 1.2 Micron Filter, Catalog # 2202-0007; BD AlarisTM Pump Infusion Set, Catalog # 2204-0007; BD AlarisTM Pump Infusion Set Yellow-Striped Microbore Tubing, Catalog # 2206-0007; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 2260-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22600-0007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22602-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22603-B007T; BD AlarisTM Pump Infusion Set Back Check Valve SmartSiteTM Y-site, Catalog # 24001-0007; BD AlarisTM Pump Infusion Set Half Set SmartSiteTM Y-site, Catalog # 2403-0007; BD AlarisTMPumpInfusionSetSmartsiteTM BagAcessNon-Vented0.2MicronFilterBondedTexiumTMClosedLowSorbingTubing(PELined)MaleLuerwithprimingcapSmartSiteTMY-site, Catalog # 24301-0007T; BD AlarisTM Pump Infusion Buerette Set Smallbore Tubing SmartsiteTM Port (Burette) 3 SmartSiteTM Y-Site, Catalog # 2441-0007; BD AlarisTM Pump Infusion Set Back Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSiteTM Y-sites, Catalog # 2450-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 24600-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing TexiumTM Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 24601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-Site, Catalog # 2465-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined), Catalog # 2466-0007.
Reason for recall
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Data sourced from openFDA. This site is unofficial and independent of the FDA.