Recall Z-0346-2018— Vyaire Medical
Class I · Terminated
Class I FDA medical device recall by Vyaire Medical, initiated 2014-11-29, terminated 2021-02-05. It names one FDA 510(k) clearance: K160764. Product: Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.. Reason: Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient..
- Recalling firm
- Vyaire Medical
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2014-11-29
- Terminated
- 2021-02-05
- Firm location
- Vernon Hills, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Reason for recall
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
Data sourced from openFDA. This site is unofficial and independent of the FDA.