Recall Z-0345-2016— DeRoyal Industries Inc

Class II · Terminated

Class II FDA medical device recall by DeRoyal Industries Inc, initiated 2015-10-14, terminated 2017-07-01. It names one FDA 510(k) clearance: K925006. Product: HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.. Reason: DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system..

Recalling firm
DeRoyal Industries Inc
Classification
Class II
Status
Terminated
Initiated
2015-10-14
Terminated
2017-07-01
Firm location
Powell, TN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.

Reason for recall

DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.