Recall Z-0340-2017— Stryker Neurovascular
Class II · Terminated
Class II FDA medical device recall by Stryker Neurovascular, initiated 2016-10-05, terminated 2017-03-23. It names 2 FDA 510(k) clearances: K010853, K980453. Product: Guider Softip XF; GUIDER/40 DEG XF/6FR/100 cm, Model number: M003101420. Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system.. Reason: Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product..
- Recalling firm
- Stryker Neurovascular
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-10-05
- Terminated
- 2017-03-23
- Firm location
- Fremont, CA, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Guider Softip XF; GUIDER/40 DEG XF/6FR/100 cm, Model number: M003101420. Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system.
Reason for recall
Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Data sourced from openFDA. This site is unofficial and independent of the FDA.