Recall Z-0337-2024— Aizu Olympus Co., Ltd.

Class II · Ongoing

Class II FDA medical device recall by Aizu Olympus Co., Ltd., initiated 2023-09-25. It names one FDA 510(k) clearance: K963033. Product: Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.. Reason: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection.

Recalling firm
Aizu Olympus Co., Ltd.
Classification
Class II
Status
Ongoing
Initiated
2023-09-25
Firm location
Aizuwakamatsu, N/A, Japan

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.

Reason for recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Data sourced from openFDA. This site is unofficial and independent of the FDA.