Recall Z-0335-2022— Osteomed, LLC

Class II · Ongoing

Class II FDA medical device recall by Osteomed, LLC, initiated 2021-10-28. It names one FDA 510(k) clearance: K971692. Product: OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Power Control Console 2 REF 450-0021-01 -Power Control Console 2 REF ECOS-450-0021-00 -OsteoMed OsteoPower Power Console 2i REF 450-0005-00 -OsteoMed D-Power Control Console 2i REF 450-0005-01 -Power Control Console 2i REF ECOS-450-0005-00 OsteoMed Footswitches: -OsteoMed BMF Footswitch REF 450-0390 -D-Bi-Directional Master Footswitch REF 450-0380 -D-Uni-Directional Footswitch II REF 450-0350 -D-Bi-Directional Footswitch II REF 450-0360 -D-Uni-Directional Master Footswitch REF 450-0370. Reason: Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient..

Recalling firm
Osteomed, LLC
Classification
Class II
Status
Ongoing
Initiated
2021-10-28
Firm location
Addison, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Power Control Console 2 REF 450-0021-01 -Power Control Console 2 REF ECOS-450-0021-00 -OsteoMed OsteoPower Power Console 2i REF 450-0005-00 -OsteoMed D-Power Control Console 2i REF 450-0005-01 -Power Control Console 2i REF ECOS-450-0005-00 OsteoMed Footswitches: -OsteoMed BMF Footswitch REF 450-0390 -D-Bi-Directional Master Footswitch REF 450-0380 -D-Uni-Directional Footswitch II REF 450-0350 -D-Bi-Directional Footswitch II REF 450-0360 -D-Uni-Directional Master Footswitch REF 450-0370

Reason for recall

Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient.