Recall Z-0329-2025— WISSNER-BOSSERHOFF GMBH
Class II · Ongoing
Class II FDA medical device recall by WISSNER-BOSSERHOFF GMBH, initiated 2022-07-11. FDA's recall record names no specific 510(k) clearance for this event. Product: SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198. Reason: When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved..
- Recalling firm
- WISSNER-BOSSERHOFF GMBH
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-07-11
- Firm location
- Wickede (Ruhr), N/A, Germany
Product description
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Reason for recall
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Data sourced from openFDA. This site is unofficial and independent of the FDA.