Recall Z-0328-2024— Aizu Olympus Co., Ltd.
Class II · Ongoing
Class II FDA medical device recall by Aizu Olympus Co., Ltd., initiated 2023-09-25. It names 2 FDA 510(k) clearances: K183525, K070983. Product: Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.. Reason: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection.
- Recalling firm
- Aizu Olympus Co., Ltd.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-09-25
- Firm location
- Aizuwakamatsu, N/A, Japan
510(k) clearances named in this recall
2 on FDA's record- K183525 — EVIS EUS Ultrasound Bronchofibervideoscope Olympus BF-UC190F
- K070983 — OLYMPUS XBF-UC180F-DT8 ULTRASONIC BRONCHOFIBERVIDEOSCOPE USED WITH ALOKA SSD-ALPHA 5/10 ULTRASOUND SYSTEM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Data sourced from openFDA. This site is unofficial and independent of the FDA.