Recall Z-0327-2024— Aizu Olympus Co., Ltd.
Class II · Ongoing
Class II FDA medical device recall by Aizu Olympus Co., Ltd., initiated 2023-09-25. It names 6 FDA 510(k) clearances: K201758, K172726, K121959 and 3 more. Product: Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT190. 510(k)s: K023984, K121959, K963033, K172726, K201758, K033225, K183419.. Reason: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection.
- Recalling firm
- Aizu Olympus Co., Ltd.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-09-25
- Firm location
- Aizuwakamatsu, N/A, Japan
510(k) clearances named in this recall
6 on FDA's record- K201758 — EVIS EXERA III Bronchovideoscope Olympus BF-XP190, EVIS EXERA III Bronchovideoscope Olympus BF-P190
- K172726 — EVIS EXERA III BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-MP190F, Single Use Biopsy Forceps FB-433D
- K121959 — EVIS EXERA III VIDEO SYSTEM
- K033225 — EVIS EXERA BRONCHOFIBERVIDEOSCOPE OLYMPUS BF TYPE XP160F, ITS ACCESSORIES AND ANCILLARY EQUIPMENT
- K023984 — MODIFICATION TO OLYMPUS BRONCHOSCOPES BF-40 SERIES, BF-240 SERIES, AND BF-160 SERIES
- K963033 — BF 240/P240/IT240 BRONCHOVIDEOSCOPE & ACCESSORIES
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT190. 510(k)s: K023984, K121959, K963033, K172726, K201758, K033225, K183419.
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Data sourced from openFDA. This site is unofficial and independent of the FDA.