Recall Z-0323-2017— Medtronic Navigation, Inc.

Class II · Terminated

Class II FDA medical device recall by Medtronic Navigation, Inc., initiated 2016-09-01, terminated 2017-03-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures.. Reason: Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field..

Recalling firm
Medtronic Navigation, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-09-01
Terminated
2017-03-27
Firm location
Louisville, CO, United States

Product description

Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures.

Reason for recall

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Data sourced from openFDA. This site is unofficial and independent of the FDA.