Recall Z-0322-2017— Medtronic Navigation, Inc.
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc., initiated 2016-09-01, terminated 2017-03-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures. Reason: Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field..
- Recalling firm
- Medtronic Navigation, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-09-01
- Terminated
- 2017-03-27
- Firm location
- Louisville, CO, United States
Product description
Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures
Reason for recall
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.
Data sourced from openFDA. This site is unofficial and independent of the FDA.