Recall Z-0319-2021— Boston Scientific Corporation

Class II · Terminated

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2020-09-22, terminated 2024-09-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application. Reason: There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line..

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Terminated
Initiated
2020-09-22
Terminated
2024-09-04
Firm location
Saint Paul, MN, United States

Product description

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

Reason for recall

There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.