Recall Z-0318-2018— Repro-Med Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Repro-Med Systems, Inc., initiated 2016-03-10, terminated 2018-02-26. It names one FDA 510(k) clearance: K122404. Product: HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.. Reason: On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification..
- Recalling firm
- Repro-Med Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-03-10
- Terminated
- 2018-02-26
- Firm location
- Chester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS22606, RMS42609 & RMS32612.
Reason for recall
On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.
Data sourced from openFDA. This site is unofficial and independent of the FDA.