Recall Z-0317-2017— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2016-09-28, terminated 2020-04-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece. Reason: Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece. An inoperable device may lead to a delay in surgery.

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-09-28
Terminated
2020-04-16
Firm location
Warsaw, IN, United States

Product description

Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece

Reason for recall

Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece. An inoperable device may lead to a delay in surgery