Recall Z-0313-2017— Ortho-Clinical Diagnostics
Class III · Terminated
Class III FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2016-06-22, terminated 2018-06-11. It names one FDA 510(k) clearance: K030067. Product: VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro qualitative detection of Hepatitis B Surface Antigen (HBsAg) in human serum and plasma (heparin, EDTA or citrate). The performance of the VITROS Immunodiagnostic Products HBsAg Controls has not been established with any other HBsAg assay.. Reason: The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2016-06-22
- Terminated
- 2018-06-11
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro qualitative detection of Hepatitis B Surface Antigen (HBsAg) in human serum and plasma (heparin, EDTA or citrate). The performance of the VITROS Immunodiagnostic Products HBsAg Controls has not been established with any other HBsAg assay.
Reason for recall
The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.