Recall Z-0312-2019— Becton Dickinson & Co.

Class II · Terminated

Class II FDA medical device recall by Becton Dickinson & Co., initiated 2018-09-24, terminated 2020-04-10. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.. Reason: A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive..

Recalling firm
Becton Dickinson & Co.
Classification
Class II
Status
Terminated
Initiated
2018-09-24
Terminated
2020-04-10
Firm location
Sparks, MD, United States

Product description

BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Reason for recall

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

Data sourced from openFDA. This site is unofficial and independent of the FDA.