Recall Z-0309-2017— Micro Therapeutics Inc, Dba Ev3 Neurovascular
Class I · Terminated
Class I FDA medical device recall by Micro Therapeutics Inc, Dba Ev3 Neurovascular, initiated 2016-10-05, terminated 2017-10-11. FDA's recall record names no specific 510(k) clearance for this event. Product: Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature. Catalog No. 105-5055V02 105-5055V02 105-5055V05 105-5065V02 105-5065V05 105-5066V02 105-5066V05. Reason: Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream..
- Recalling firm
- Micro Therapeutics Inc, Dba Ev3 Neurovascular
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2016-10-05
- Terminated
- 2017-10-11
- Firm location
- Irvine, CA, United States
Product description
Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature. Catalog No. 105-5055V02 105-5055V02 105-5055V05 105-5065V02 105-5065V05 105-5066V02 105-5066V05
Reason for recall
Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Data sourced from openFDA. This site is unofficial and independent of the FDA.